PURE WATER GRADE 2
             PURE WATER GRADE 3

 

QUALITY STANDARD

ISO 3696 
ASTM 
NCCLS 
EP/USP 

EL. SPEC. CONDUCIBILITY  25°C (D) 0.2-3 MicroS/cm
EL. SPEC. RESISTIVITY 25°C (D2) 10-18,2 MOhm.cm

RO CUBIC/D 

grade 3
Type III 
Type III 
conform (w/filter 0,2mcr) 

RO CUBIC/D2

grade 2
Type II
Type II
conform

 

 

A different model for all the user need
The RO CUBIC line provides the ability to adapt each unit to the specific needs of the user. 
Indeed RO CUBIC is managed by an electronic control with PLC software and able to manage all functions and active control of the machine, so it is possible from the user request specific changes that are best suited to your needs.

 

3 possible flow.

- RO 15 CUBIC 15 lt/h
- RO 30 CUBIC 15+15 lt/h (differentiable flow)
- RO 60 CUBIC 60 lt/h

 
3 purity grade.

- RO CUBIC purified water
- RO CUBIC/d pure water GRADE 3
- RO CUBIC/d2 pure water GRADE 2


water feed request


- TDS max 
- HD max 
- Iron max 
- Manganese max 
- clhor max 
- Total bacteria  max 
- SDI 

1000 ppm
35 °F
100 ppb
5 ppb
0,1 ppm
5 UFC/ml
< 5
4 purification stages.

- sediment microfilter 
- reverse osmosis with booster pump
- mixed bed ion excange resins (models /d and/d2)
- anti bacteria microfilter 0,2 micron (model /d2)


4 mode of operation.

- ON/OFF with operator
- automatic ON/OFF with 2 sensors level for tank (min/max)
- ON with operator and automatico OFF (with sensor level)
- ON with operator and OFF with liter meter (volume set)

 

TECHNICAL FEATURES

APPLICATIONS


- min. feed pressure 
- max. feed pressure 
- range of temp.water to be treat. 
- ELECTRIC POWER 
- Hydraulic connections 
 

- max operative pressure 

bar 1,5
bar 4,0
°C 5-35
V 220 ac
IN pipe PE 10
OUT pipe PE 6/4
DRAIN pipe PE 6/4
bar 8,0

HPLC chromatography
Spectro-photometric analysis under atomic absorption
Reagents preparation/dilution
Colorimetric and qualitative analysis
Feeding of ultra-pure water systems, surge tanks, etc.
Feeding of glass works washers
Feeding of steam generators
 

 

PRINT pag of catalogue pag.18-19    [DEMINERALIZERS CATALOGUE]     [menu]     [home]     [contact]